Package: masscor 0.0.7.1

masscor: Mass Measurement Corrections

Mass measurement corrections and uncertainties using calibration data, as recommended by EURAMET's guideline No. 18 (2015) ISBN:978-3-942992-40-4 . The package provides classes, functions, and methods for storing information contained in calibration certificates and converting balance readings to both conventional mass and real mass. For the latter, the Magnitude of the Air Buoyancy Correction factor employs models (such as the CIMP-2007 formula revised by Picard, Davis, Gläser, and Fujii (2008) <doi:10.1088/0026-1394/45/2/004>) to estimate the local air density using measured environmental conditions.

Authors:Cristhian Paredes [aut, cre]

masscor_0.0.7.1.tar.gz
masscor_0.0.7.1.zip(r-4.7)masscor_0.0.7.1.zip(r-4.6)masscor_0.0.7.1.zip(r-4.5)
masscor_0.0.7.1.tgz(r-4.6-any)masscor_0.0.7.1.tgz(r-4.5-any)
masscor_0.0.7.1.tar.gz(r-4.7-any)masscor_0.0.7.1.tar.gz(r-4.6-any)
masscor_0.0.7.1.tgz(r-4.6-emscripten)
manual.pdf |manual.html
card.svg |card.png
masscor/json (API)
NEWS

# Install 'masscor' in R:
install.packages('masscor', repos = c('https://crparedes.r-universe.dev', 'https://cloud.r-project.org'))

Bug tracker:https://github.com/crparedes/masscor/issues

Datasets:
  • Box.E2.MS.Kit - Calibration data of a weights kit in the interval 1 mg to 1 kg
  • E2.MS.20g - Calibration data of a E2 class mass standard of nominal mass 20 g.
  • minimalCert - Minimal 'calibCert()' object
  • MT.XP.2002 - Calibration data for a balance Mettler Toledo XP 2002
  • MT.XPE.204 - Calibration data for a balance Mettler Toledo XPE 204

On CRAN:

Conda:

3.74 score 11 scripts 237 downloads 14 exports 5 dependencies

Last updated from:8e72bd5b54. Checks:9 OK. Indexed: yes.

TargetResultTimeFilesSyslog
linux-devel-x86_64OK125
source / vignettesOK161
linux-release-x86_64OK120
macos-release-arm64OK80
macos-oldrel-arm64OK72
windows-develOK71
windows-releaseOK75
windows-oldrelOK84
wasm-releaseOK109

Exports:airDensityairDensityHASLcalibCertconvertMassUnitsSIconvMassMABCmassStandardmassStandardKitnormalizedErroruncertAirDensityuncertConvMassuncertErrorCorruncertMABCuncertReading

Dependencies:DEoptimRMASSmetRologynumDerivrobustbase

masscor package

Rendered frommasscor.Rmdusingknitr::rmarkdownon May 18 2026.

Last update: 2021-09-10
Started: 2021-06-11

Readme and manuals

Help Manual

Help pageTopics
'masscor': Mass Measurement Corrections and Uncertainties.masscor-package
Models for calculating air density based on environmental conditionsairDensity
Air density estimation using only height above sea level.airDensityHASL
Calibration data of a weights kit in the interval 1 mg to 1 kgBox.E2.MS.Kit
Information of balance calibration certificatecalibCert
Conversion between mass units of The International System of UnitsconvertMassUnitsSI
Corrects a balance reading using balance calibration data.convMass
Calibration data of a E2 class mass standard of nominal mass 20 g.E2.MS.20g
Magnitude of the Air Buoyancy CorrectionMABC
Creates an object of class '"massStandard"'.massStandard
Creates an object of class '"massStandardKit"'.massStandardKit
Minimal 'calibCert()' objectminimalCert
Calibration data for a balance Mettler Toledo XP 2002MT.XP.2002
Calibration data for a balance Mettler Toledo XPE 204MT.XPE.204
Calculates normalized in balance verification using a mass standardnormalizedError
S3 method for plotting objects of class '"calibCert"'plot.calibCert
S3 method for printing objects of class '"calibCert"'print.calibCert
S3 method for printing objects of class '"massStandard"'print.massStandard
S3 method for printing objects of class '"massStandardKit"'print.massStandardKit
Uncertainties in air density calculationsuncertAirDensity
Uncertainty in conventional mass valueuncertConvMass
Uncertainty due to mass correction using calibration certificateuncertErrorCorr
Uncertainty of the Magnitude of the Air Buoyancy Correction factoruncertMABC
Uncertainty of balance readingsuncertReading